GE Analytical Instruments ©2010
124 of 226
DLM 74001-04 Rev. A
Chapter 6: Calibration and System Protocols
Accuracy, Precision, and Calibration Verification
The Sievers 500 RL On-Line TOC Analyzer automatically calculates accuracy, precision, and calibration
verification values. Use this procedure to confirm that the Analyzer’s current calibration is accurate.
GE Analytical Instruments recommends performing a verification after replacement of consumables, such as
the pump heads (with pump tubing), DI water cartridge, or UV lamp, and after calibration. If your configuration
does not measure conductivity, you will not be using the conductivity standard for this protocol.
To perform accuracy, precision, and calibration verification, follow this procedure:
1. If the Analyzer is taking measurements, press the
Stop Analysis
button.
2. Press the
Menu
button, select the
Maintenance
tab, press the
Cal/Ver/Validate
button, and press the
Acc/Prec/Ver
button.
1. If the Analyzer is taking measurements, press the
Stop Analysis
button.
2. If you have a Super
iOS
System, the
Select Rinse
screen appears. Otherwise, go to step 10.
3. Select one of the following rinse options, and then press the
Next
button.
• No Rinse
• Before
• After
• Before and After
4. If you have selected
No Rinse
or
After
, and there are vials in the Super
iOS
, remove them now. Press
Next
to continue. Wait for the Analyzer to drain each of the Super
iOS
vial chambers (about 2 minutes).
5. If you have selected to run a rinse, the Analyzer will prompt you to do one of the following:
• Insert the Super
iOS
Rinse vial set into the Super
iOS
.
• Or, if using individual rinse vials, press
Next
for additional instructions. Insert the individual vials in
the Super
iOS
, and then press
No Set
to proceed.
6. Do one of the following:
• If you have a purchased vial set, insert the
Accuracy, Precision, an Calibration Verification
standards
cartridge into the Super
iOS
System with the label facing away from the Analyzer and press
Next
.
• If you have purchased individual vials rather than a vial set in a cartridge, confirm the value shown
on the Conductivity Standard label. Next, make sure the vials are inserted into the
iOS
System vial
ports in the following order:
Port 1 = empty
Port 2 = Verification Blank
Port 3 = Accuracy/Precision and Verification Standard
Port 4 = Conductivity Standard
Press the
Next
button, and then press the
No Set
button.
Содержание Sievers 500 RL
Страница 8: ...GE Analytical Instruments 2010 8 of 226 DLM 74001 04 Rev A ...
Страница 10: ...GE Analytical Instruments 2010 10 of 226 DLM 74001 04 Rev A ...
Страница 36: ...GE Analytical Instruments 2010 36 of 220 DLM 74001 04 Rev A ...
Страница 66: ...GE Analytical Instruments 2010 66 of 226 DLM 74001 04 Rev A Chapter 3 Installation ...
Страница 152: ...GE Analytical Instruments 2010 152 of 226 DLM 74001 04 Rev A Chapter 7 Maintenance ...
Страница 170: ...GE Analytical Instruments 2010 170 of 226 DLM 74001 04 Rev A Chapter 8 Troubleshooting ...
Страница 177: ...Appendix A GE Analytical Instruments 2010 177 of 186 DLM 74001 04 Rev A Figure 51 Left Side Analyzer Dimensions ...
Страница 178: ...Appendix A GE Analytical Instruments 2010 178 of 186 DLM 74001 04 Rev A ...
Страница 185: ...Notes GE Analytical Instruments 2010 185 of 186 DLM 74001 04 Rev A 186 ...
Страница 186: ...Notes GE Analytical Instruments 2010 186 of 186 DLM 74001 04 Rev A 186 ...