GE Analytical Instruments ©2010
121 of 226
DLM 74001-04 Rev. A
Chapter 6: Calibration and System Protocols
If the label on the Conductivity Standard shows a different value,
press
Edit
, enter the value, press
Enter
, and then press
Next
to continue.
When prompted by the Analyzer, remove the first four vials and insert the next four vials into the
Super
iOS
System vial ports in the following order:
Port 1 = empty
Port 2 = TOC Calibration Blank
Port 3 = TOC Calibration Standard (1.0 ppm KHP)
Port 4 = TOC Calibration Standard (500 ppb KHP)
Press the
Next
button. Proceed to Step 12.
9. If you have a standard
iOS
System:
• Open the door to the
iOS
System and wait 30 seconds for water to drain.
• Insert the first TOC Calibration Blank into the
iOS
System and press
Next
.
• When prompted, remove the TOC Calibration Blank from the
iOS
System, insert the second TOC
Calibration Blank, and press
Next
.
• When prompted, remove the TOC Calibration Blank from the
iOS
System, insert the TOC Calibration
Standard (1.50 ppm KHP), and press
Next
.
• When prompted, remove the TOC Calibration Standard (1.50 ppm KHP) from the
iOS
System.
Confirm the value of the Conductivity Standard and insert the vial into the
iOS
System. If the label
shows a value of 25 µS/cm HCl, press
Next
to continue. If the label shows a different value, press
Edit
, enter the value, press
Enter
, and then press
Next
. If your configuration does not measure
conductivity, you will not be prompted to enter a conductivity value and you can skip this step.
• When prompted, remove the Conductivity Standard from the
iOS
System, insert the third TOC
Calibration Blank, and press
Next
.
• When prompted, remove the TOC Calibration Blank from the
iOS
System, insert the TOC Calibration
Standard (1.0 ppm KHP), and press
Next
.
• When prompted, remove the TOC Calibration Standard (1.0 ppm KHP) from the
iOS
System, insert
the TOC Calibration Standard (500 ppb KHP), and press
Next
.
10. After the last standard has been analyzed, the Calibration Summary screen is displayed. Part 1 of the
summary screen displays the data for the first two blanks and the 1.50 ppm standard (see Figure 26).
Press the
Next
button.
11. Part 2 of the summary displays the data for the conductivity standard (see Figure 27). Press the
Next
button.
12. Part 3 of the summary displays data for the blank, 1.0 ppm standard, and 500 ppb standard (see
Figure 27).
13. Press the
button if you want to print the Calibration results.
Содержание Sievers 500 RL
Страница 8: ...GE Analytical Instruments 2010 8 of 226 DLM 74001 04 Rev A ...
Страница 10: ...GE Analytical Instruments 2010 10 of 226 DLM 74001 04 Rev A ...
Страница 36: ...GE Analytical Instruments 2010 36 of 220 DLM 74001 04 Rev A ...
Страница 66: ...GE Analytical Instruments 2010 66 of 226 DLM 74001 04 Rev A Chapter 3 Installation ...
Страница 152: ...GE Analytical Instruments 2010 152 of 226 DLM 74001 04 Rev A Chapter 7 Maintenance ...
Страница 170: ...GE Analytical Instruments 2010 170 of 226 DLM 74001 04 Rev A Chapter 8 Troubleshooting ...
Страница 177: ...Appendix A GE Analytical Instruments 2010 177 of 186 DLM 74001 04 Rev A Figure 51 Left Side Analyzer Dimensions ...
Страница 178: ...Appendix A GE Analytical Instruments 2010 178 of 186 DLM 74001 04 Rev A ...
Страница 185: ...Notes GE Analytical Instruments 2010 185 of 186 DLM 74001 04 Rev A 186 ...
Страница 186: ...Notes GE Analytical Instruments 2010 186 of 186 DLM 74001 04 Rev A 186 ...