Chapter 4: Basic Analyzer Operation
Adjusting Analog Output Values
Adjust the 4-20 mA output values by following these steps:
1. Open the Analyzer door and connect wiring from one of the 4-20 mA outputs on TB 3 to a digital
multimeter.
2. Select the
I/O
tab.
3. Press the
4-20mA Outputs
button.
4. Press the button that corresponds to the output to which you connected in Step 1 (either
Analog Out 1
,
Analog Out 2
, or
Analog Out 3
).
5. Press the
Adj. AnalogOut
button.
6. Press the
4 mA
button.
7. Specify a value between 0 and 65535 using the keypad and press the
Enter
button.
8. Check the value that is output to the multimeter. If the reading is not 4 mA
0.01 mA, go back to Step 6
and adjust the value. Repeat until the multimeter measures 4 mA
0.01 mA.
9. Press the
20 mA
button.
10. Specify a value between 0 and 65535 using the keypad and press the
Enter
button.
11. Check the value that is output to the multimeter. If the reading is not 20 mA
0.01 mA, go back to Step
9 and adjust the value. Repeat until the multimeter measures 20 mA
0.01 mA.
12. Disconnect the multimeter wires from TB3 and close the Analyzer door.
Setting Up Serial Output
To download data from the Analyzer directly to a computer, you must set the
Baud Rate
for the serial (RS-232)
port. The default is
9600
.
1. Select the
I/O
tab.
2. Press the
Serial
button.
3. Press the
Baud Rate
button. Select
9600
,
19200
, or
38400
and press the
Enter
button to save the value,
or press
Cancel
to retain the current setting without making any changes. Consult your computer
manual to determine the maximum baud rate for your computer’s serial port. The default value is
9600
.
4. Press the
Output Format
button. If your data acquisition system is set up to work with Sievers 400 ES
TOC Analyzers and you want to use a serial output format compatible with the older format, press the
400 ES
button. Otherwise,
500 RL
should be selected as the output format. The default format is
500 RL
.
Содержание Sievers 500 RL
Страница 8: ...GE Analytical Instruments 2010 8 of 226 DLM 74001 04 Rev A ...
Страница 10: ...GE Analytical Instruments 2010 10 of 226 DLM 74001 04 Rev A ...
Страница 36: ...GE Analytical Instruments 2010 36 of 220 DLM 74001 04 Rev A ...
Страница 66: ...GE Analytical Instruments 2010 66 of 226 DLM 74001 04 Rev A Chapter 3 Installation ...
Страница 152: ...GE Analytical Instruments 2010 152 of 226 DLM 74001 04 Rev A Chapter 7 Maintenance ...
Страница 170: ...GE Analytical Instruments 2010 170 of 226 DLM 74001 04 Rev A Chapter 8 Troubleshooting ...
Страница 177: ...Appendix A GE Analytical Instruments 2010 177 of 186 DLM 74001 04 Rev A Figure 51 Left Side Analyzer Dimensions ...
Страница 178: ...Appendix A GE Analytical Instruments 2010 178 of 186 DLM 74001 04 Rev A ...
Страница 185: ...Notes GE Analytical Instruments 2010 185 of 186 DLM 74001 04 Rev A 186 ...
Страница 186: ...Notes GE Analytical Instruments 2010 186 of 186 DLM 74001 04 Rev A 186 ...