GE Analytical Instruments ©2010
108 of 226
DLM 74001-04 Rev. A
Chapter 5: Password Protection and DataGuard
Configuring Login Settings
To further ensure Analyzer security, DataGuard allows you to configure login settings as follows:
1. Select the
DataGuard
tab.
2. Press the
Setup
button.
• Press the
Password Expires
button to set the number of days for which a Password is valid. The
default value is
90
days.
• Press the
Min Length
button to set the minimum character length for a Password. The default value
is
3
characters.
• Press the
Login Timeout
button to set the number of minutes of allowable keypad inactivity before
users are required to re-enter their login information. The default value is
5
minutes.
• Press the
Password Config
button select the Password Protection menu for the Analyzer. Select
either
Cal Lockout
or
Menu Lockout
. When the Analyzer is set to
Cal Lockout
, only operations that
can affect measurements are password protected. When the Analyzer is set to
Menu Lockout
, all
menus are password protected.
Changing User Passwords
All users can change their own Passwords by following these steps:
1. Log in with a valid User ID and Password.
2. Select the
DataGuard
tab.
3. Press the
Change Password
button.
4. Enter the old Password and press
Enter
.
5. Specify a new Password and press
Enter
.
6. Verify the new Password and press
Enter
.
Administrators can change the Password associated with any User ID by pressing the
Password
button on the
Edit User
screen (“Editing User Information” on page 107).
Dealing with Forgotten Passwords
If users forget their Passwords, the administrator must assign new Passwords by following these steps:
1. Log in to the Analyzer with an administrator-level User ID and Password.
2. Select the
DataGuard
tab.
3. Press the
Edit User
button.
Содержание Sievers 500 RL
Страница 8: ...GE Analytical Instruments 2010 8 of 226 DLM 74001 04 Rev A ...
Страница 10: ...GE Analytical Instruments 2010 10 of 226 DLM 74001 04 Rev A ...
Страница 36: ...GE Analytical Instruments 2010 36 of 220 DLM 74001 04 Rev A ...
Страница 66: ...GE Analytical Instruments 2010 66 of 226 DLM 74001 04 Rev A Chapter 3 Installation ...
Страница 152: ...GE Analytical Instruments 2010 152 of 226 DLM 74001 04 Rev A Chapter 7 Maintenance ...
Страница 170: ...GE Analytical Instruments 2010 170 of 226 DLM 74001 04 Rev A Chapter 8 Troubleshooting ...
Страница 177: ...Appendix A GE Analytical Instruments 2010 177 of 186 DLM 74001 04 Rev A Figure 51 Left Side Analyzer Dimensions ...
Страница 178: ...Appendix A GE Analytical Instruments 2010 178 of 186 DLM 74001 04 Rev A ...
Страница 185: ...Notes GE Analytical Instruments 2010 185 of 186 DLM 74001 04 Rev A 186 ...
Страница 186: ...Notes GE Analytical Instruments 2010 186 of 186 DLM 74001 04 Rev A 186 ...