Chapter 6. Calibration and System Protocols
Overview
The Analyzer is calibrated at the factory and should require recalibration only once per year. When replacing
items that affect analysis, such as the UV lamp or pump heads, perform a verification of the calibration; only re-
calibrate if verification indicates that a new calibration is necessary.
The Sievers 500 RL On-Line TOC Analyzer can be calibrated using one of two methods: single-point calibration at
1.5 ppm; or multi-point calibration at 500 ppb, 1.0 ppm, and 1.5 ppm. GE Analytical Instruments recommends
single-point calibration for most applications. Multi-point calibration is available as an alternative procedure for
customers.
The Sievers 500 RL On-Line TOC Analyzer facilitates the calibration and verification process by handling the
necessary calculations internally; you will have the opportunity to accept or reject the calibration during the
procedure. Calibration and verification are accomplished using Sievers Standards in 40-mL and 30-mL vials,
sampled via the Analyzer’s
iOS
System.
If your Analyzer has a standard
iOS
System (one vial port), you will need to insert the appropriate standard into
the
iOS
when prompted by the Analyzer. If your Analyzer has a Super
iOS
System (four vial ports), you can insert
all standards required for most system protocols into the
iOS
at one time; when running multi-point calibration,
you will need to switch vials once during the procedure.
If your Analyzer is equipped with a sample inlet block, you must install an IOS to calibrate and verify the Analyzer.
Note:
Only a single-point calibration
or
a multi-point calibration needs to be performed. Do
not perform both types of calibration.
Содержание Sievers 500 RL
Страница 8: ...GE Analytical Instruments 2010 8 of 226 DLM 74001 04 Rev A ...
Страница 10: ...GE Analytical Instruments 2010 10 of 226 DLM 74001 04 Rev A ...
Страница 36: ...GE Analytical Instruments 2010 36 of 220 DLM 74001 04 Rev A ...
Страница 66: ...GE Analytical Instruments 2010 66 of 226 DLM 74001 04 Rev A Chapter 3 Installation ...
Страница 152: ...GE Analytical Instruments 2010 152 of 226 DLM 74001 04 Rev A Chapter 7 Maintenance ...
Страница 170: ...GE Analytical Instruments 2010 170 of 226 DLM 74001 04 Rev A Chapter 8 Troubleshooting ...
Страница 177: ...Appendix A GE Analytical Instruments 2010 177 of 186 DLM 74001 04 Rev A Figure 51 Left Side Analyzer Dimensions ...
Страница 178: ...Appendix A GE Analytical Instruments 2010 178 of 186 DLM 74001 04 Rev A ...
Страница 185: ...Notes GE Analytical Instruments 2010 185 of 186 DLM 74001 04 Rev A 186 ...
Страница 186: ...Notes GE Analytical Instruments 2010 186 of 186 DLM 74001 04 Rev A 186 ...