GE Analytical Instruments ©2010
157 of 226
DLM 74001-04 Rev. A
Chapter 8: Troubleshooting
before continuing with analysis. You may need to contact Technical Support to confirm that the Analyzer is
functioning properly.
51 - History Flash Error
Message: “A problem has been detected in the history data flash. Please contact GE.”
Explanation: The Analyzer experienced a problem when trying to store data in the flash memory. It is possible a
section of the flash memory is damaged. To protect against losing data, the system will automatically abort
analysis. Contact Technical Support.
52 - User List Error
Message: “The user list is full. Archive the user list before adding more users.”
Explanation: DataGuard allows a maximum of 100 users. This error displays when 100 user accounts have been
created. The Analyzer will abort analysis. If you have inactive accounts in the user list, you can archive users
(
DataGuard
tab
Archive Users
) to remove them from the list; you will then be able to add additional users.
53 - History Data Error
Message: “The history data is full. Archive the history data before taking more measurements.”
Explanation: This error displays when the data history is completely filled. Export (
Data
tab
Export
) or Print
(
Data
tab
the data) before proceeding.
54 - Audit Trail Error
Message: “The audit trail is full. Archive the audit trail before taking more measurements.”
Explanation: This error displays when the audit trail is completely filled. To comply with 21 CFR Part 11, you must
either Export (
DataGuard
tab
Export AuditTrail
) or Print (
DataGuard
tab
Print Audit Trail
) the audit trail
before proceeding.
55 - Linearity Vial Error
Message: “The Linearity vials appear to be in the wrong order. Check the vials to assure they are in the correct
order.”
Explanation: This error generally displays when the vials in a Linearity protocol are not inserted into the Super
iOS System in the correct order. Consult Chapter 6: Calibration and System Protocols for the correct vial order.
56 - Sample Pump Error
Message: “The sample pump is not pumping. Please refer to the Oper. and Maint. Manual. Contact GE is problem
persists.”
Explanation: The Analyzer has detected that the sample pump is not turning and thus not drawing sample water
through the Analyzer. Try cycling power to the Analyzer. If you recently performed maintenance on the Analyzer,
confirm that all connections were plugged in properly. Contact technical support for additional assistance.
58 - UV Lamp Error
Message: “A problem has been detected with the UV lamp. Inspect the cable. Refer to the Oper. and Maint.
Manual for recommendations.”
Explanation: The Analyzer has detected that the UV lamp is not functioning properly. Try cycling power to the
Analyzer. If you recently performed maintenance on the Analyzer, confirm that the UV lamp power supply is
properly connected and that the lamp is secured to the UV reactor housing. Also confirm that the glass window
in the UV reactor shows the lamp is illuminated when the Analyzer is powered on. Contact technical support for
additional assistance.
59 - Language Error
Message: “Incorrect language pack installed. Refer to the Oper. and Maint. Manual for recommendations.”
Содержание Sievers 500 RL
Страница 8: ...GE Analytical Instruments 2010 8 of 226 DLM 74001 04 Rev A ...
Страница 10: ...GE Analytical Instruments 2010 10 of 226 DLM 74001 04 Rev A ...
Страница 36: ...GE Analytical Instruments 2010 36 of 220 DLM 74001 04 Rev A ...
Страница 66: ...GE Analytical Instruments 2010 66 of 226 DLM 74001 04 Rev A Chapter 3 Installation ...
Страница 152: ...GE Analytical Instruments 2010 152 of 226 DLM 74001 04 Rev A Chapter 7 Maintenance ...
Страница 170: ...GE Analytical Instruments 2010 170 of 226 DLM 74001 04 Rev A Chapter 8 Troubleshooting ...
Страница 177: ...Appendix A GE Analytical Instruments 2010 177 of 186 DLM 74001 04 Rev A Figure 51 Left Side Analyzer Dimensions ...
Страница 178: ...Appendix A GE Analytical Instruments 2010 178 of 186 DLM 74001 04 Rev A ...
Страница 185: ...Notes GE Analytical Instruments 2010 185 of 186 DLM 74001 04 Rev A 186 ...
Страница 186: ...Notes GE Analytical Instruments 2010 186 of 186 DLM 74001 04 Rev A 186 ...