Chapter 8: Troubleshooting
Step 2: Visual Inspection
After confirming that there are no warning or error messages indicating a specific problem, proceed to a visual
inspection of the Analyzer to identify any obvious physical problems or symptoms. Begin with an external visual
inspection, and proceed to an internal visual inspection. If possible, perform the inspection with the Analyzer in
analysis mode so you can verify that relevant modules are performing as usual (described below).
External Inspection
•
Confirm that there are no leaks present at the sample inlet or the waste connections.
•
Confirm that all cables are securely attached to the Analyzer.
•
If you are sampling from vials, confirm that the needle in the
iOS
System is not obstructed.
Internal Inspection
Open the Analyzer front panel and do the following:
•
Check for leaks throughout the Analyzer; there should not be moisture on the inside floor of the
Analyzer case.
•
Confirm that the sample pump is turning.
•
Check to make sure there are no leaks from tubing in the DI water system.
•
Confirm that the DI water cartridge is full. If necessary, add water up to the
Full
line.
•
If maintenance work has recently been performed on the Analyzer, confirm that electronics cables,
such as the connector between the analog and ID boards, are securely attached.
•
Make sure there are no air bubbles present in any of the DI loop tubing.
•
Make sure tubing is not bent or obstructed.
•
Make sure the UV lamp is working. During analysis, the window on the UV reactor will illuminate when
the lamp is functioning properly.
Note that if tubing has become bent, kinked, or otherwise damaged, a tubing replacement kit is available from
GE Analytical Instruments Technical Support.
Step 3: Review Solutions for Basic Problems
If a visual inspection of the Analyzer does not help you identify the source of the problem you are experiencing,
consult the specific problems discussed in this section.
Содержание Sievers 500 RL
Страница 8: ...GE Analytical Instruments 2010 8 of 226 DLM 74001 04 Rev A ...
Страница 10: ...GE Analytical Instruments 2010 10 of 226 DLM 74001 04 Rev A ...
Страница 36: ...GE Analytical Instruments 2010 36 of 220 DLM 74001 04 Rev A ...
Страница 66: ...GE Analytical Instruments 2010 66 of 226 DLM 74001 04 Rev A Chapter 3 Installation ...
Страница 152: ...GE Analytical Instruments 2010 152 of 226 DLM 74001 04 Rev A Chapter 7 Maintenance ...
Страница 170: ...GE Analytical Instruments 2010 170 of 226 DLM 74001 04 Rev A Chapter 8 Troubleshooting ...
Страница 177: ...Appendix A GE Analytical Instruments 2010 177 of 186 DLM 74001 04 Rev A Figure 51 Left Side Analyzer Dimensions ...
Страница 178: ...Appendix A GE Analytical Instruments 2010 178 of 186 DLM 74001 04 Rev A ...
Страница 185: ...Notes GE Analytical Instruments 2010 185 of 186 DLM 74001 04 Rev A 186 ...
Страница 186: ...Notes GE Analytical Instruments 2010 186 of 186 DLM 74001 04 Rev A 186 ...