Commissioning
Proline Prosonic Flow 93 PROFIBUS DP/PA
92
Hauser
6.3.2
Quick Setup "Commissioning"
a0008715-en
Fig. 85:
Quick Setup "Commissioning"
+
+
+
E
E
0424
0425
0423
0402
0422
1002
B
2000
Esc
E
+
-
XXX.XXX.XX
HOME-POSITION
n
o
m
p
q
Unit
Volume Flow
Unit
Temperature
Unit
Viscosity
Unit
Length
Unit
Velocity
QS
Commission.
Language
Presetting
Quick Setup
System Units
Volume
Viscosity
Length
Velocity
Quit
Configure another unit?
NO
YES
Selection system units
Autom. configuration display?
Another quick setup?
NO
NO
Automatical parameterization
of the display
YES
Temperature
Deliver Settings
y
Selection pre-settings
Actual Settings
!
Note!
• The display returns to the cell SETUP COMMISSIONING (1002) if you press the
X
key combination during
parameter interrogation. The stored parameters remain valid.
• The "Commissioning" Quick Setup must be carried out before one of the Quick Setups explained below is run.
m
The "DELIVERY SETTINGS" option sets every selected unit to the factory setting.
The "ACTUAL SETTINGS" accepts the units you configured beforehand.
n
Only units not yet configured in the current Quick Setup are offered for selection in each cycle. The volume unit is
derived from the volume flow unit.
o
The "YES" option remains visible until all the units have been configured.
"NO" is the only option displayed when no further units are available.
p
The "automatic parameterization of the display" option contains the following basic settings/factory settings
YES
Main line = mass flow
Additional line = totalizer 1
Information line = operating/system condition
NO
The existing (selected) settings remain.
q
The execution of other Quick Setups is described in the following sections.