Troubleshooting
Proline Prosonic Flow 93 PROFIBUS DP/PA
140
Hauser
121
S: A / C COMPATIB.
!: # 121
0x0F
BAD
Device Failure
Constant
Amplifier and I/O
board only partially
compatible
Cause:
Due to different software versions, I/O board and
amplifier board are only partially compatible
(possibly restricted functionality).
!
Note!
• This message is only listed in the error history.
• Nothing is displayed on the display.
Remedy:
Module with lower software version has either to
be updated by FieldCare with the required software
version or the module has to be replaced.
205
S: LOAD T-DAT
!: # 205
0x0F
BAD
Device Failure
Constant
Save to T-DAT
failed
Cause:
Data backup (download) to T-DAT failed, or error
when accessing (uploading) the calibration values
stored in the T-DAT.
Remedy:
1.
Check whether the T-DAT is correctly plugged
into the amplifier board. Using the spare part
set number, check whether the new,
replacement DAT is compatible with the
existing measuring electronics.
2.
Replace the T-DAT if it is defective.
Check that the new, replacement DAT is
compatible with the measuring electronics.
Check the:
- Spare part set number
- Hardware revision code
3.
Replace measuring electronics boards if
necessary.
4.
Plug the T-DAT into the amplifier board.
206
S: SAVE T-DAT
!: # 206
0x0F
BAD
Device Failure
Constant
Restore from
T-DAT failed
261
S: COMMUNICAT. I/O
$
: # 261
0x18
0x19
0x1A
BAD
No Commu-
nication
O.K.
Low
High
Communication
failure
Cause:
Communication error. No data reception between
amplifier and I/O board or faulty internal data
transfer.
Remedy:
Check whether the electronics boards are correctly
inserted in their holders
392
S: SIGNAL LOW CH1
$
: # 392
0x0F
BAD
Device Failure
Constant
Attenuation of
acoustic
measurement
section too high
Cause:
System error. Attenuation of acoustic measurement
section too high.
Remedy:
– Check to see if the coupling fluid must be
renewed.
– It is possible that the fluid indicates too much
attenuation.
– It is possible that the pipe indicates too much
attenuation.
– Check the sensor spacing (Installation
dimensions).
– Reduce the number of traverses if possible.
393
S: SIGNAL LOW CH1
$
: # 393
No.
Device status message
(local display)
PROFIBUS measured value status
Extended
diagnostic
message in the
PROFIBUS
master
Cause/remedy (spare part Page 145 ff.)
Q
u
a
lit
y
c
o
d
e
(
H
EX
)
Me
as
ure
d
val
u
e
s
ta
tus
Qu
a
lit
y
s
ta
tu
s
Qu
a
lit
y
s
u
b
st
a
tu
s
Li
m
its
Содержание Proline Prosonic Flow 93
Страница 2: ......
Страница 168: ...www endress com worldwide BA00076D 06 EN 13 10 71121236 FM SGML6 0 ProMoDo...