Wiring
Proline Prosonic Flow 93 PROFIBUS DP/PA
68
Hauser
4.3.3
PROFIBUS DP connection diagram
Permanent assignment board (order version 93***-***********J)
A0014363
Fig. 68:
Connecting the transmitter, cable cross-section max. 2.5 mm
2
(AWG 14)
a
Connection compartment cover
b
Cable for power supply: 85 to 260 V AC, 20 to 55 V AC, 16 to 62 V DC
Terminal No. 1: L1 for AC, L+ for DC
Terminal No. 2: N for AC, L
−
for DC
c
Ground terminal for protective ground
d
Fieldbus cable:
Terminal No. 26: B (RxD/TxD-P)
Terminal No. 27: A (RxD/TxD-N)
e
Ground terminal for fieldbus cable shield
Observe the following:
– the shielding and grounding of the fieldbus cable
– that the stripped and twisted lengths of cable shield to the ground terminal are as short as possible
f
Service adapter for connecting service interface FXA193 (Fieldcheck, FieldCare)
g
Power supply cable for external bus terminator (optional):
Terminal No. 24: +5 V
Terminal No. 25: DGND
A (RxD/TxD-N)
B (RxD/TxD-P)
1
2
c
e
f
b
b
d
d
g
22 23
20 21
24 25 26 27
a
DGND
+5 V
g
L1 (L+)
N (L–)