Proline Prosonic Flow 93 PROFIBUS DP/PA
Wiring
Hauser
65
4.2.2
Grounding Prosonic Flow P DN 15 to 65 (½ to 2½")
The Prosonic Flow P DN 15 to 65 (½ to 2½") is grounded via the cable gland.
Procedure
1.
Guide the connecting cables through the cable gland.
2.
Position the connecting cables in such a way that the heat shrink tube ends at the edge of the
cable gland (the stripped part of the connecting cable is thus in the correct position).
3.
Tighten the nut of the cable gland (this twists the pins in the cable gland against the shielding
of the connecting cable and establishes grounding).
4.2.3
Cable specification for connecting cable
Only use the connecting cables supplied by Hauser.
The connecting cables are available in different lengths
132 ff.
For the cable specifications, see
Operation in areas with strong electrical interference
The measuring system complies with the general safety requirements in accordance with
EN 61010, the EMC requirements of IEC/EN 61326 "Emission as per Class A requirements" and
NAMUR Recommendation NE 21.