Proline Prosonic Flow 93 PROFIBUS DP/PA
Commissioning
Hauser
101
BASIC FUNCTION (G)
→
SYSTEM PARAMETER (GLA)
→
CONFIGURATION (660)
→
MEASURING MODE (6601)
→
Selection of one of the following options:
– UNIDIRECTIONAL (factory setting) = only the positive flow portions
– BIDIRECTIONAL = the positive and negative flow components
7.
Configuration of totalizers 1 to 3:
The measuring device has three totalizers. The following example describes the configuration
of the totalizer using totalizer 1 as an example.
• Using the CHANNEL function (6133), you can determine the measured variable (e.g. mass
flow) to be transmitted to the PROFIBUS master (Class 1) as a totalizer value:
a. Select BASIC FUNCTION (G)
→
PROFIBUS PA (GCA)
→
TOTALIZER (613)
→
SELECT TOTALIZER (6130).
b. Select the option TOTALIZER 1.
c. Go to the function CHANNEL (6133).
d. Select the option VOLUME FLOW.
Possible settings
→
see following table.
• Enter the desired unit for the totalizer:
BASIC FUNCTION (G)
→
PROFIBUS PA (GCA)
→
TOTALIZER (613)
→
UNIT TOTALIZER (6134)
• Configure totalizer status (e.g. totalize):
BASIC FUNCTION (G)
→
PROFIBUS PA (GCA)
→
TOTALIZER (613)
→
SET TOTALIZER (6135)
→
Select the option TOTALIZE
• Set the totalizer mode:
BASIC FUNCTION (G)
→
PROFIBUS PA (GCA)
→
TOTALIZER (613)
→
TOTALIZER MODE (6137)
→
Selection of one of the following options:
– BALANCE (factory setting): calculates the positive and negative flow portions
– POSITIVE: calculates the positive flow portions
– NEGATIVE: calculates the negative flow portions
– HOLD VALUE: The totalizer remains at the last value
!
Note!
For the calculation of the positive and negative flow portions (BALANCE) or the negative
flow portions (NEGATIVE) to be carried out correctly, the selection BIDIRECTIONAL must
be active in the function BASIC FUNCTION (G)
→
SYSTEM PARAMETER (GLA)
→
CONFIGURATION (660)
→
MEASURING MODE (6601).
Possible settings
Totalizer value/measured variable (channel 1 = active)
ID for CHANNEL function
VOLUME FLOW CHANNEL 1 (factory setting totalizer 1 to 3)
273
OFF
0
Totalizer value/measured variable (channel 1 + channel 2 =
active)
ID for CHANNEL function
VOLUME FLOW CHANNEL 1 (factory setting totalizer 1 to 3)
273
VOLUME FLOW CHANNEL 2
529
AVERAGE VOLUME FLOW
567
VOLUME FLOW SUM
568
VOLUME FLOW DIFFERENCE
569
OFF
0
!
Note!
If, when performing the PROFIBUS network configuration, the module or the function TOTAL was integrated in
slot 9, 10 or 11, the measured variable selected in the CHANNEL function for the totalizers 1 to 3 is transmitted
cyclically to the PROFIBUS master (Class 1).