Proline Prosonic Flow 93 PROFIBUS DP/PA
Commissioning
Hauser
91
!
Note!
The display returns to the cell SETUP SENSOR (1001) if you press the
X
key combination during parameter
interrogation. The stored parameters remain valid.
m
The choice of system units only influences the functions UNIT TEMPERATURE (0422),
UNIT LENGTH (0424) and UNIT VELOCITY (0425).
n
If a channel is selected for which a Quick Setup has already been executed, the previous values are overwritten.
o
During each run, all the options can be selected. If settings were made during a previous run, these are overwritten.
p
If YES is selected, the value measured during Quick Setup is accepted in the appropriate function. If NO is selected,
the measurement is discarded and the original value remains.
q
The POSITION SENSOR function (6884) only appears if the CLAMP ON option is selected in the MESURE function
and the number of traverses selected in the SENSOR CONFIGURATION function (6882) is 2 or 4.
r
The WIRE LENGTH function (6885) only appears if the CLAMP ON option is selected in the MESURE function and
the number of traverses selected in the SENSOR CONFIGURATION function (6882) is 1 or 3.
s
The ARC LENGTH function (6887) only appears if the INSERTION option is selected in the MESURE function and
the DUAL PATH option is selected in the SENSOR CONFIGURATION function (6882).