Proline Prosonic Flow 93 PROFIBUS DP/PA
Troubleshooting
Hauser
139
041
S: TRANSM. HW-DAT
$
: # 041
0x0F
BAD
Device Failure
Constant
T-DAT failure
Cause
:
1.
T-DAT is not plugged into the amplifier board
correctly (or is missing).
2.
T-DAT is defective.
Remedy:
1.
Check whether the T-DAT is correctly plugged
into the amplifier board.
2.
Replace the T-DAT if it is defective.
Check that the new, replacement DAT is
compatible with the measuring electronics.
Check the:
- Spare part set number
- Hardware revision code
3.
Replace measuring electronics boards if
necessary.
4.
Plug the T-DAT into the amplifier board.
042
S: TRANSM. SW-DAT
$
: # 042
0x0F
BAD
Device Failure
Constant
T-DAT data
inconsistent
082
S: SENS. DOWN CH1
$
: # 082
0x13
BAD
Sensor Failure
Constant
Interruption
between sensor
and transmitter
CH1
Cause:
System error. Connection between sensor channel
1/2 and transmitter interrupted.
Remedy:
– Check the cable connection between the sensor
and the transmitter.
– Check that the sensor connector is fully screwed
in.
– The sensor may be defective.
– Incorrect sensor connected.
Change the sensor type specified:
Access:
BASIC FUNCTION
→
SENSOR DATA
→
SENSOR
PARAMETER
→
SENSOR TYPE
083
S: SENS. DOWN CH2
$
: # 083
0x13
BAD
Sensor Failure
Constant
Interruption
between sensor
and transmitter
CH2
085
S: SENSOR UP CH1
$
: # 085
0x13
BAD
Sensor Failure
Constant
Interruption
between sensor
and transmitter
CH1
Cause:
System error. Connection between sensor channel
1/2 and transmitter interrupted.
Remedy:
– Check the cable connection between the sensor
and the transmitter.
– Check that the sensor connector is fully screwed
in.
– The sensor may be defective.
– Incorrect sensor connected.
Change the sensor type specified:
Access:
BASIC FUNCTION
→
SENSOR DATA
→
SENSOR
PARAMETER
→
SENSOR TYPE
086
S: SENSOR UP CH2
$
: # 086
0x13
BAD
Sensor Failure
Constant
Interruption
between sensor
and transmitter
CH2
No.
Device status message
(local display)
PROFIBUS measured value status
Extended
diagnostic
message in the
PROFIBUS
master
Cause/remedy (spare part Page 145 ff.)
Q
u
a
lit
y
c
o
d
e
(
H
EX
)
Me
as
ure
d
val
u
e
s
ta
tus
Qu
a
lit
y
s
ta
tu
s
Qu
a
lit
y
s
u
b
st
a
tu
s
Li
m
its
Содержание Proline Prosonic Flow 93
Страница 2: ......
Страница 168: ...www endress com worldwide BA00076D 06 EN 13 10 71121236 FM SGML6 0 ProMoDo...