2/Theory of Operation
1503-0151-000
3/27/97
2-11
Power Entry Module
Toroid
ac to dc rectifier
•
Type
Line Filter with ac Connector & 5 X 20
mm Fuse holder
•
Maximum leakage current
< 5
µ
A @ 250V/50Hz
•
Fuses
2A 250V. Slow blowing 5X20 mm
•
Approvals
UL 1283, CSA C22.2/8, VDE 0565 &
EN 60320
•
Type
High efficiency toroid
•
Power capacity
54.5 VA
•
Voltage rating
85-270Vac / 14-45 Vac
•
Leakage current
25
µ
A Max.
•
Power efficiency
85% - 90% @ 115Vac 60Hz, 60VA
•
Safety screen thickness
0.005" Cu.
•
Isolation hipot voltage
4KV.
•
Thermal fuse
Thermik S01 110
o
C
•
Dimension
115 mm diameter x 58 mm thickness
•
Weight
4.8 lb
•
Mounting method
Epoxy center hole 0.5 in. ID.
•
Topology
Full wave bridge/voltage doubling with
auto/voltage adapter
•
Rectifier type
GBU8D 8A, 200V
•
Peak current surge
300A (JEDEC Method).
•
Power efficiency
92% - 95%
•
Output filter capacitor
2 x 5600uF 63WVdc, 3.93A allowable
ripple current at 120 Hz, 85
o
C, 2000 hr.
load life at 85
o
C 50V 3.93A 120 Hz rip
•
Output ripple voltage
< 2.5 Vp-p
•
dc output voltage range
24 - 62 Vdc
•
Auto-voltage adapting threshold
140/155 Vac
Содержание Aestiva 7900 SmartVent
Страница 2: ...1503 0151 000 3 27 97 ...
Страница 4: ...1503 0151 000 3 27 97 ...
Страница 12: ...Notes viii 1503 0151 000 3 27 97 ...
Страница 18: ...1 Introduction 1 6 1503 0151 000 5 26 0 Notes ...
Страница 44: ...2 Theory of Operation 2 26 1503 0151 000 3 27 97 Notes ...
Страница 46: ...3 Post Service Checkout 3 2 1503 0151 000 5 26 0 Notes ...
Страница 158: ...8 Schematics 8 2 7900 Ventilator 1503 0151 000 1 97 Figure 8 1 System connection block diagram ...
Страница 159: ...8 Schematics 7900 Ventilator 1503 0151 000 1 97 8 3 Figure 8 2 SIB schematic diagram interface cable connections ...
Страница 160: ...8 Schematics 8 4 7900 Ventilator 1503 0151 000 1 97 Figure 8 3 Patient interface cable Machine side ...