6/ Repair Procedures
1503-0151-000
5/26/0
6-45
5. Remove the Tabletop from the Modulus SE. See Modulus SE Service Manual
section 5.3.
6. Disconnect the four color coded tubing connectors from the SIB Assembly
tubing.
7. Carefully cut and remove the cable ties that secure the electrical harness to the
pan.
8. Disconnect connector J1 from P1 on the SIB Assembly.
9. Open the Vanity panel on the rear of the Anesthesia Machine.
10. While supporting the Interface Assembly, remove the two M4 keps nuts (item
5, Figure 6-37) using a 7 mm nutdriver.
11. Carefully route the harness and tubing through the pan to remove the
assembly.
1.
Label, patient interface, 1001-3464-000
2.
Harness SA, SIB, 1503-3081-000
3.
Panel, sensor interface, Modulus SE, 1004-3933-000
4.
Spacer, sensor interface panel, 1004-3934-000 (2)
5.
Nut, keps M4, 0144-3717-314 (2)
Figure AA.85.083
Patient Interface/Harness Assembly (shown mounted on the Modulus SE left
side)
AA.85083
1
2
3
4
5
Содержание Aestiva 7900 SmartVent
Страница 2: ...1503 0151 000 3 27 97 ...
Страница 4: ...1503 0151 000 3 27 97 ...
Страница 12: ...Notes viii 1503 0151 000 3 27 97 ...
Страница 18: ...1 Introduction 1 6 1503 0151 000 5 26 0 Notes ...
Страница 44: ...2 Theory of Operation 2 26 1503 0151 000 3 27 97 Notes ...
Страница 46: ...3 Post Service Checkout 3 2 1503 0151 000 5 26 0 Notes ...
Страница 158: ...8 Schematics 8 2 7900 Ventilator 1503 0151 000 1 97 Figure 8 1 System connection block diagram ...
Страница 159: ...8 Schematics 7900 Ventilator 1503 0151 000 1 97 8 3 Figure 8 2 SIB schematic diagram interface cable connections ...
Страница 160: ...8 Schematics 8 4 7900 Ventilator 1503 0151 000 1 97 Figure 8 3 Patient interface cable Machine side ...