2/Theory of Operation
1503-0151-000
3/27/97
2-3
Safety Features
•
Dual redundant airway over pressure protection, linked to the Pmax
setting
•
Volume over-delivery limits and protection
•
Ohmeda proprietary hose connections - fixed manifolds to reduce
leaks
•
Uses proven mechanical components
•
10 VA - Oxygen compartment separation
•
500 psi burst protection
2.2 Mechanical
Figure 2-3
Supply gas inlet, filtered
Supply Gas
which can be powered from O2 or Air, is supplied from the anesthesia
machine at a pressure of 241 to 690 kPa (35 to 100 psi). This supply gas is filtered
through the 5 micron filter and water trap to further eliminate any minute particles
of contaminate. The filter does not significantly reduce the output pressure on the
downstream side of the filter.
Figure 2-4
Gas Inlet Valve
Inlet
Filter
Supply
Gas
Hose
35 - 100 psi
Supply inlet
Outlet
Inlet
Filter
Gas
Inlet
Valve
(GIV)
35 - 100 psi
Supply
Gas
Hose
35 - 100 psi
Supply inlet
Outlet
Содержание Aestiva 7900 SmartVent
Страница 2: ...1503 0151 000 3 27 97 ...
Страница 4: ...1503 0151 000 3 27 97 ...
Страница 12: ...Notes viii 1503 0151 000 3 27 97 ...
Страница 18: ...1 Introduction 1 6 1503 0151 000 5 26 0 Notes ...
Страница 44: ...2 Theory of Operation 2 26 1503 0151 000 3 27 97 Notes ...
Страница 46: ...3 Post Service Checkout 3 2 1503 0151 000 5 26 0 Notes ...
Страница 158: ...8 Schematics 8 2 7900 Ventilator 1503 0151 000 1 97 Figure 8 1 System connection block diagram ...
Страница 159: ...8 Schematics 7900 Ventilator 1503 0151 000 1 97 8 3 Figure 8 2 SIB schematic diagram interface cable connections ...
Страница 160: ...8 Schematics 8 4 7900 Ventilator 1503 0151 000 1 97 Figure 8 3 Patient interface cable Machine side ...