4/Tests and Troubleshooting
4-14
1503-0151-000
5/26/0
Pressure Limit Switch FAIL Instructions:
Go to the Discrete I/O signal menu and check the DPL (Drive Pressure Limit)
switch status. Disconnect pressure switch tubing from manifold and apply 104
±
4
cmH2O to activate switch and verify switch status. If signal status does not agree
with switch status, disconnect J8 on microcontroller board and measure resistance
to determine function of switch.