2/Theory of Operation
2-4
1503-0151-000
3/27/97
During normal operation the
GIV (Gas Inlet Valve)
is open to allow the supply gas
to flow to the ventilator manifold. This valve provides a shut off of the supply gas
to the ventilator when the ventilator is not in use. The GIV also shuts off supply
gas to the ventilator under failure conditions as the microprocessor or over-pres-
sure switch demands. The output from the GIV remains at the filtered supply gas
pressure.
Figure 2-5
Non-Relieving Pressure Regulator
The pressure regulator is a
Non-relieving Pressure Regulator
which regulates
the high pressure filtered supply gas, oxygen or medical air, down to 172 kPa
(25 psi).
Figure 2-6
Flow Control Valve
The Flow Control Valve is controlled by the microcontroller. Signals are sent to the
Flow Control valve which correspond to the generated flow requirements called for
by ventilator settings and sensor signals. The Flow Control Valve modulates the
incoming 25 psi drive gases to a variable output from 0 to 120 liters per minute at
pressures nominally ranging from minus 10 to plus 100 cmH
2
O.
Pressure
Regulator
Inlet
Filter
Gas
Inlet
Valve
(GIV)
25 psi
Supply
Gas
Hose
35 - 100 psi
Supply inlet
Outlet
Flow
Valve
Pressure
Regulator
Inlet
Filter
Gas
Inlet
Valve
(GIV)
Supply
Gas
Hose
35 - 100 psi
Supply inlet
25 psi
0 - 120 Lpm
Содержание Aestiva 7900 SmartVent
Страница 2: ...1503 0151 000 3 27 97 ...
Страница 4: ...1503 0151 000 3 27 97 ...
Страница 12: ...Notes viii 1503 0151 000 3 27 97 ...
Страница 18: ...1 Introduction 1 6 1503 0151 000 5 26 0 Notes ...
Страница 44: ...2 Theory of Operation 2 26 1503 0151 000 3 27 97 Notes ...
Страница 46: ...3 Post Service Checkout 3 2 1503 0151 000 5 26 0 Notes ...
Страница 158: ...8 Schematics 8 2 7900 Ventilator 1503 0151 000 1 97 Figure 8 1 System connection block diagram ...
Страница 159: ...8 Schematics 7900 Ventilator 1503 0151 000 1 97 8 3 Figure 8 2 SIB schematic diagram interface cable connections ...
Страница 160: ...8 Schematics 8 4 7900 Ventilator 1503 0151 000 1 97 Figure 8 3 Patient interface cable Machine side ...