6/ Repair Procedures
6-30
1503-0151-000
5/26/0
Removing the flow control valve
1. Disconnect the cable connector (J7) from microcontroller board.
Figure 6-25
Disconnect the flow control valve cable
2. Use the 3 mm hex wrench to loosen the two (2) captive mounting screws and
lift out the flow control valve. (See figure 6-26)
3. Examine the seat and two O-rings for damage; replace as necessary.