2/Theory of Operation
1503-0151-000
3/27/97
2-7
Figure 2-9
MOBO (Mechanical Over pressure Bleed Off)
The Mechanical Over pressure Bleed Off Valve
is a mechanical valve operating
even if there is no electrical power. The MOBO has two functions. First, it serves
as a third level of redundancy to the ventilator's pressure limit control functions,
providing pressure relief at approximately 110 cmH
2
O. Second, the MOBO serves
as a backup in the event of a complete blockage of the exhalation valve system,
relieving circuit pressure at approximately 30 cmH
2
O under such failure conditions.
Mechanical
Overpressure
Bleed Off
(MOBO)
Drive Gas Check Valve
(DGCV)
Supply
Gas
Hose
Manifold
Pressure
Transducer
Flow
Valve
Pressure
Regulator
Inlet
Filter
Gas
Inlet
Valve
(GIV)
25 psi
25 mm port
17 mm port
Exhaust
to Scavenging
System
Exhaust to
Ambient
Exhalation
Manifold
35 - 100 psi
Supply inlet
Содержание Aestiva 7900 SmartVent
Страница 2: ...1503 0151 000 3 27 97 ...
Страница 4: ...1503 0151 000 3 27 97 ...
Страница 12: ...Notes viii 1503 0151 000 3 27 97 ...
Страница 18: ...1 Introduction 1 6 1503 0151 000 5 26 0 Notes ...
Страница 44: ...2 Theory of Operation 2 26 1503 0151 000 3 27 97 Notes ...
Страница 46: ...3 Post Service Checkout 3 2 1503 0151 000 5 26 0 Notes ...
Страница 158: ...8 Schematics 8 2 7900 Ventilator 1503 0151 000 1 97 Figure 8 1 System connection block diagram ...
Страница 159: ...8 Schematics 7900 Ventilator 1503 0151 000 1 97 8 3 Figure 8 2 SIB schematic diagram interface cable connections ...
Страница 160: ...8 Schematics 8 4 7900 Ventilator 1503 0151 000 1 97 Figure 8 3 Patient interface cable Machine side ...