6/ Repair Procedures
1503-0151-000
5/26/0
6-3
3. Remove the dual hose for drive/return gas connecting the bellows to the exha-
lation valve and remove the exhaust hose from the exhalation valve manifold.
1.
Track on which the ventilator mounts in the Excel/Modulus SE Gas Machine
2.
Slot into which the locking screw fits for secure mounting
3.
Locking screw, holds the ventilator in place on the track when tightened
4.
Rear panel locking tabs, push toward the center to release cover
5.
Rear panel of the Excel/Modulus SE.
Figure 6-2
Removing the ventilator from an Excel/Modulus SE Anesthesia Gas Machine
4. Unplug the ventilator power cord from the outlet box and thread it out through
the opening made by cover removal.
5. Disconnect the gas supply hose from the gas machine power outlet and thread
it out through the opening made by panel removal.
6. Loosen the securing knob under the ventilator on the right hand side.
7. Pull the ventilator back so it is flush with the back of the gas machine housing.
8. Remove the cable from the ventilator to the SIB board and the RS232 serial
port connection if applicable.
CAUTION:
w
w
w
w
The ventilator weighs approximately 40 pounds. Use care in its removal.
9. See Section 6.3, service shelf setup
Содержание Aestiva 7900 SmartVent
Страница 2: ...1503 0151 000 3 27 97 ...
Страница 4: ...1503 0151 000 3 27 97 ...
Страница 12: ...Notes viii 1503 0151 000 3 27 97 ...
Страница 18: ...1 Introduction 1 6 1503 0151 000 5 26 0 Notes ...
Страница 44: ...2 Theory of Operation 2 26 1503 0151 000 3 27 97 Notes ...
Страница 46: ...3 Post Service Checkout 3 2 1503 0151 000 5 26 0 Notes ...
Страница 158: ...8 Schematics 8 2 7900 Ventilator 1503 0151 000 1 97 Figure 8 1 System connection block diagram ...
Страница 159: ...8 Schematics 7900 Ventilator 1503 0151 000 1 97 8 3 Figure 8 2 SIB schematic diagram interface cable connections ...
Страница 160: ...8 Schematics 8 4 7900 Ventilator 1503 0151 000 1 97 Figure 8 3 Patient interface cable Machine side ...