6/ Repair Procedures
1503-0151-000
5/26/0
6-7
6.3. Setting up the Service Shelf
The service shelf folds for easy storage. When setting up ensure the four locking
screws are secure.
CAUTION:
w
The circuit boards are electrostatic sensitive. Set the 7900 service shelf
on an anti-static work station and wear a wrist grounding strap while
working on the boards.
1.
Notch for locking screw on the ventilator
2.
Spare locking knob
3.
Locking knobs (4). Hold the service shelf in position when unfolded to hold the
ventilator while servicing
Figure 6-6
Setting up the service shelf
1. Ensure the service shelf is setup on the anesthesia machine's work surface.
CAUTION:
w
The ventilator weighs approximately 40 pounds. Use care in its removal.