4/Tests and Troubleshooting
1503-0151-000
5/26/0
4-1
4.1. Overview
WARNING:
w
w
w
w
Do not perform testing or maintenance on medical instruments while
they are being used on a patient; patient injury may result.
This section includes pass and fail tests with calibration procedures for 7900 Ven-
tilator control module. Operator calibration information is covered in the 7900 Ven-
tilator O&M manual.
Testing Requirements
The 7900 Ventilator, an integral part of the Ohmeda Anesthesia System, can be
tested while the control module is mounted in the anesthesia machine. From the
test results, you must determine if it is necessary to remove the control module to
gain access for service and/or maintenance.
The ventilator must be removed from the anesthesia system and placed on a spe-
cial service shelf in order to gain access for service, maintenance and parts re-
placement. This service shelf allows you to set the ventilator on the work surface
of the anesthesia system and reconnect hoses and cables.
If you are performing routine maintenance it is necessary to complete the entire
section. It is always necessary to perform section "3/Post-Service Checkout" after
this section.
WARNING:
w
w
w
w
Post-Service Checkout is required after you complete this section. You
must perform section " 3/Post-Service Checkout" after performing any
maintenance, service or repair. Failure to do so may result in patient injury.
The 7900 Ventilator has a self test built into its software. This test determines that
the software is functioning properly and that the electronic circuits on the circuit
boards are functional.
The following list of the tests run at power-up:
•
*Sequential watchdog
•
*Logical watchdog
•
*Data RAM walking pattern test
•
*FLASH ROM CRC verification
•
Gas inlet valve test (electrical and pneumatic)
•
Calibration of the manifold sensor
If the gas inlet valve causes the self test to fail on power up an alarm occurs and is
displayed appropriately "Gas Inlet Valve Circuit Failure" Also an error code
(1002C) will register on the error code service mode menu.
Содержание Aestiva 7900 SmartVent
Страница 2: ...1503 0151 000 3 27 97 ...
Страница 4: ...1503 0151 000 3 27 97 ...
Страница 12: ...Notes viii 1503 0151 000 3 27 97 ...
Страница 18: ...1 Introduction 1 6 1503 0151 000 5 26 0 Notes ...
Страница 44: ...2 Theory of Operation 2 26 1503 0151 000 3 27 97 Notes ...
Страница 46: ...3 Post Service Checkout 3 2 1503 0151 000 5 26 0 Notes ...
Страница 158: ...8 Schematics 8 2 7900 Ventilator 1503 0151 000 1 97 Figure 8 1 System connection block diagram ...
Страница 159: ...8 Schematics 7900 Ventilator 1503 0151 000 1 97 8 3 Figure 8 2 SIB schematic diagram interface cable connections ...
Страница 160: ...8 Schematics 8 4 7900 Ventilator 1503 0151 000 1 97 Figure 8 3 Patient interface cable Machine side ...