6/ Repair Procedures
1503-0151-000
5/26/0
6-13
1.
Power supply board has two captive securing screws. Finger tighten, low torque.
2
Power supply board.
3.
Power supply board connector for isolation transformer cable.
4.
Key hole slots for angle mounting flange.
5.
Battery connection (J1) to power supply board. See step 5 in previous proce-
dure.
Figure 6-11
Removing the power supply board