4/Tests and Troubleshooting
4-20
1503-0151-000
5/26/0
Error Code
Description of error logged
Service Action
10031
Minimum system monitoring
caused by flow sensor EEPROM
failure
1. Refer to Section 4 Display A/D Channels,
Test Flow Sensors. Replace faulty flow sen-
sors
2. Check cable connections at SIB and the
SIB to 7900 cable
3. Replace the SIB
4. Replace the CPU board
10032
Minimum system monitoring
caused by DCGO (French)
1. System has DCGO in aux..position but
language not set to French
2. Check Discrete I/O Signal menu for sta-
tus
3. Check SIB cables
4. Replace CPU board
10035
Non-volatile memory EEPROM
data out of range
Reboot the system, if problem continues
replace the CPU board
10036
Non-volatile memory EEPROM
region protected value error
Reboot the system, if problem continues
replace the CPU board
10038
1 of 3 safety relevant parameter
areas has difference in SRC tripli-
cate storage (external RAM prob-
lem)
Reboot the system, if problem continues
replace the CPU board
10039
All 3 safety relevant parameter
areas are different in SRC storage
(external RAM problem)
Reboot the system, if problem continues
replace the CPU board
1003A
There are no more EEPROM
areas to switch to (EEPROM prob-
lem)
Reboot the system, if problem continues
replace the CPU board
1003D
Minimum system monitoring
caused by system parameter fail-
ure (external RAM problem)
Reboot the system, if problem continues
replace the CPU board
1003E
Minimum system monitoring
caused by manifold pressure sen-
sor failure
Reboot the system, if problem continues
replace the CPU board
10042
The system clock has been repro-
grammed (CPU timer problem)
Reboot the system, if problem continues
replace the CPU board
10045
Sustained airway pressure hold-
ing for more than 10 seconds
No Service Action Required
10046
Minimum system shutdown low
battery, less than 11.35V while
system is running on battery
1. Charge the battery
2. If the unit has been plugged in for 12
hours and the problem continues, replace
battery
Содержание Aestiva 7900 SmartVent
Страница 2: ...1503 0151 000 3 27 97 ...
Страница 4: ...1503 0151 000 3 27 97 ...
Страница 12: ...Notes viii 1503 0151 000 3 27 97 ...
Страница 18: ...1 Introduction 1 6 1503 0151 000 5 26 0 Notes ...
Страница 44: ...2 Theory of Operation 2 26 1503 0151 000 3 27 97 Notes ...
Страница 46: ...3 Post Service Checkout 3 2 1503 0151 000 5 26 0 Notes ...
Страница 158: ...8 Schematics 8 2 7900 Ventilator 1503 0151 000 1 97 Figure 8 1 System connection block diagram ...
Страница 159: ...8 Schematics 7900 Ventilator 1503 0151 000 1 97 8 3 Figure 8 2 SIB schematic diagram interface cable connections ...
Страница 160: ...8 Schematics 8 4 7900 Ventilator 1503 0151 000 1 97 Figure 8 3 Patient interface cable Machine side ...