6/ Repair Procedures
6-2
1503-0151-000
5/26/0
6.2. Removing the 7900 ventilator from the
Ohmeda Excel/Modulus SE Gas Machine
There are two versions of 7900 ventilator/anesthesia machine mounting, Integrat-
ed and non-integrated.
1.
Ventilator
2.
Ventilator shroud
3.
Middle shelf
Figure 6-1
Excel 210 with mid-shelf and integrated 7900 ventilator
Integrated;
the ventilator fits in a shroud which has a latching back cover.
Removing the 7900 ventilator from the integrated
machine
Remove the rear panel from the gas machine ventilator shroud.
1. Push the locking clips on the panel toward the center.
2. Allow the panel to hang from its retaining cord.
CAUTION:
w
Do not pull the exhalation valve manifold hoses straight down from the
block assembly to remove them. Too much force is needed if you pull
straight down. This added force can cause the hoses to release suddenly
and possibly cause damage to the valve block or cause injury to the per-
son removing the hose. To remove the hoses, use a careful, light twist-
ing, back and forth movement while you apply pressure downward.
Содержание Aestiva 7900 SmartVent
Страница 2: ...1503 0151 000 3 27 97 ...
Страница 4: ...1503 0151 000 3 27 97 ...
Страница 12: ...Notes viii 1503 0151 000 3 27 97 ...
Страница 18: ...1 Introduction 1 6 1503 0151 000 5 26 0 Notes ...
Страница 44: ...2 Theory of Operation 2 26 1503 0151 000 3 27 97 Notes ...
Страница 46: ...3 Post Service Checkout 3 2 1503 0151 000 5 26 0 Notes ...
Страница 158: ...8 Schematics 8 2 7900 Ventilator 1503 0151 000 1 97 Figure 8 1 System connection block diagram ...
Страница 159: ...8 Schematics 7900 Ventilator 1503 0151 000 1 97 8 3 Figure 8 2 SIB schematic diagram interface cable connections ...
Страница 160: ...8 Schematics 8 4 7900 Ventilator 1503 0151 000 1 97 Figure 8 3 Patient interface cable Machine side ...