4/Tests and Troubleshooting
1503-0151-000
5/26/0
4-33
Remarks
The calibration is performed as follows:
1. Open the flow valve to the starting point that was found (where
manifold pressure is 91-92 cm H
2
O).
2. Wait for the flow to stabilize.
3. Once the flow stabilized, set the flow to 12 L/min.
4. Wait for the flow to stabilize.
5. Perform steps 1-6 replacing step 3 with 11 L/min., 10 L/min., 9 L/min.
1 L/min..
Therefore, there are 12 points that are calculated for the bleed resistor calibration.
Calibration points may be greater than the listed cm H
2
O, but if any one of these
points is less than the cmH
2
O listed below, calibration fails. If the calibration pass-
es, the calculated bleed resistor calibration points are stored in EEPROM.
67 cm H
2
0
59 cm H
2
0
52 cm H
2
0
45 cm H
2
0
38 cm H
2
0
30 cm H
2
0
24 cm H
2
0
18 cm H
2
0
12 cm H
2
0
7 cm H
2
0
-5 cm H
2
0
-5 cm H
2
0
Содержание Aestiva 7900 SmartVent
Страница 2: ...1503 0151 000 3 27 97 ...
Страница 4: ...1503 0151 000 3 27 97 ...
Страница 12: ...Notes viii 1503 0151 000 3 27 97 ...
Страница 18: ...1 Introduction 1 6 1503 0151 000 5 26 0 Notes ...
Страница 44: ...2 Theory of Operation 2 26 1503 0151 000 3 27 97 Notes ...
Страница 46: ...3 Post Service Checkout 3 2 1503 0151 000 5 26 0 Notes ...
Страница 158: ...8 Schematics 8 2 7900 Ventilator 1503 0151 000 1 97 Figure 8 1 System connection block diagram ...
Страница 159: ...8 Schematics 7900 Ventilator 1503 0151 000 1 97 8 3 Figure 8 2 SIB schematic diagram interface cable connections ...
Страница 160: ...8 Schematics 8 4 7900 Ventilator 1503 0151 000 1 97 Figure 8 3 Patient interface cable Machine side ...