Repair
12-36
Service Manual
Figure 12-20.
Antennae PCB and UFL Connectors Removal
To replace the Antennae PCBs
1.
Gently thread wires through from the front to the rear of the Main PCB, with the
ferrite bead remaining visible and each antenna PCB slipping through the left or
right opening on the Main PCB.
2.
Gently connect both golden UFL antennae connectors onto the SBC PCB.
3.
Wire tie ferrite bead to the Main PCB, capturing right antenna PCB wire under-
neath the wire tie, while ensuring the wire remains free under the Main PCB.
4.
Slide the left-hand antenna PCB under the supporting clip and onto its standouts.
5.
Slip right-hand antenna PCB back in under the plastic LCD clamp plate.
6.
Replace the Main PCB. Reference
7.
Reassemble the monitoring system chassis. Reference
Содержание Nellcor GR101704
Страница 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Страница 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Страница 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Страница 96: ...Operation 6 34 Service Manual a Press an alternate menu option b Press EXIT AND RESTART to exit SERVICE MODE...
Страница 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Страница 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Страница 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Страница 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Страница 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Страница 235: ......