Repair
12-28
Service Manual
7.
Place the fan back down in its enclosure.
8.
Replace the battery cradle in the back chassis. Reference
9.
Reassemble the monitoring system chassis. Reference
Battery Interconnect PCB Replacement
To remove the Battery Interconnect PCB
1.
Follow all previous steps for removing the battery cradle. Reference
, p. 12-25. It is not necessary to remove the power supply PCB.
2.
Disconnect the five-pin J2 connector from the recess between the power supply
PCB and the wire-tied cable assembly with the ferrite bead.
3.
Rotate the battery cradle to ensure battery interconnect PCB accessibility.
4.
Remove both T-10 torx screws.
5.
Lift from the battery cradle.
Содержание Nellcor GR101704
Страница 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Страница 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Страница 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Страница 96: ...Operation 6 34 Service Manual a Press an alternate menu option b Press EXIT AND RESTART to exit SERVICE MODE...
Страница 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Страница 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Страница 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Страница 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Страница 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Страница 235: ......