Power Components Replacement
Service Manual
12-25
8.
Reverse the process for replacement.
Battery Cradle Replacement
Removing the Battery Cradle and Components
1.
Disassemble the monitoring system chassis. Reference
2.
Move the front chassis off to the side.
3.
Remove all four T-10 torx screws, two on top, and two deep underneath, using a
long-reach driver.
Figure 12-14.
Battery Cradle Removal
4.
Lift out the battery cradle, supporting both the battery cradle and still connected wires.
Note:
Cooling fan is still attached by wires, but is free-floating.
5.
Set it on a static-free work surface.
Caution:
Ensure the work surface is clean and free of debris.
6.
Remove the Power Supply PCB. Reference
7.
Remove the Battery Interconnect PCB. Reference
Содержание Nellcor GR101704
Страница 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Страница 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Страница 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Страница 96: ...Operation 6 34 Service Manual a Press an alternate menu option b Press EXIT AND RESTART to exit SERVICE MODE...
Страница 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Страница 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Страница 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Страница 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Страница 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Страница 235: ......