Service Mode
Service Manual
6-31
a.
Press SAVE CHANGES to retain the change.
b.
Press CANCEL to leave the default as it was.
Restore Factory Defaults
Proceed only if institutional defaults require replacement with the standard
factory defaults.
To restore institutional defaults back to factory defaults
1.
Access the service function menu. Reference
To access service function menus
2.
Select SETTINGS.
3.
Select RESTORE FACTORY DEFAULTS.
4.
Select the desired option.
a.
Press YES to restore the factory defaults.
b.
Press CANCEL to retain the factory defaults.
5.
Select the desired option.
a.
Press BACK to return to the previous submenu.
b.
Press EXIT MENU.
6.5.2
Service Menu
, p. 10-44, for monitoring screen cal-
ibration instructions. Reference
Software and Firmware Upgrades
for software or firmware upgrade instructions.
6.5.3
Logs Menu
Error, event, and software update logs remain in flash until the monitoring
system overwrites them and will survive both power cycles and software
upgrades. If a user chooses to “clear” a log, the monitoring system renames
the current log files, rather than deleting them. Possible scenarios for either file
corruption or a partial loss of data in a log would include backdating the mon-
itoring system when it has already recorded data to an existing date or a power
interrupt during retrieval of a log.
Содержание Nellcor GR101704
Страница 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Страница 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Страница 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Страница 96: ...Operation 6 34 Service Manual a Press an alternate menu option b Press EXIT AND RESTART to exit SERVICE MODE...
Страница 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Страница 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Страница 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Страница 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Страница 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Страница 235: ......