![background image](http://html1.mh-extra.com/html/covidien/nellcor-gr101704/nellcor-gr101704_service-manual_2673898164.webp)
Modification and Testing
10-42
Service Manual
10.7
Test Data Sheet Form
Job #
Model
Name
NBRPMS
Serial
Number
Software
Version
Table 10-6.
Performance and Functional Tests
Item
Results
Remarks
Performance Tests
Battery check
Pass
Fail
Power up defaults
•
Power-on self-test (POST)
Pass
Fail
•
Alarms limits
Pass
Fail
•
Alarm management: SatSeconds,
SPD, and pulse rate delay
Pass Fail
Monitoring settings
Pass
Fail
Operational setup
Pass
Fail
Nurse call
Pass
Fail
Monitoring screen calibration
Pass
Fail
Overall Performance Check
•
Sensor LED excitation test
Pass
Fail
•
Operation on a live subject
Pass
Fail
Functional Tests
BPM test
Pass
Fail
73-77%
57-63 bpm
73-77%
194-206 bpm
73-77%
57-63 bpm
SpO
2
test
Pass
Fail
73-77%
57-63 bpm
88-92%
57-63 bpm
73-77%
57-63 bpm
Modulation test
Pass
Fail
Light level test
Pass
Fail
DATE: TESTS
PERFORMED
BY:
Содержание Nellcor GR101704
Страница 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Страница 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Страница 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Страница 96: ...Operation 6 34 Service Manual a Press an alternate menu option b Press EXIT AND RESTART to exit SERVICE MODE...
Страница 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Страница 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Страница 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Страница 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Страница 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Страница 235: ......