Repair
12-32
Service Manual
6.
Remove the cables connected to the five-pin J23 receptacle, the four-pin J22
receptacle, and the three-pin J2 receptacle on the Main PCB.
7.
Cut the wire tie securing the ferrite bead to the wiring harness.
To replace the left power cable assembly
1.
Properly orient the new left power cable assembly.
2.
Tighten a wire tie around the ferrite bead to secure it to the wiring harness.
3.
Firmly connect the five-pin Mini-mate black cable connector to the J23 beige
receptacle on the Main PCB.
4.
Firmly connect the four-pin red and black (DC) cable and connector to the J22
white receptacle on the Main PCB.
5.
Firmly connect the three-pin black fan cable and connector to the J2 black recep-
tacle on the Main PCB.
6.
Firmly connect the five-pin Mini-mate black cable and connector to its J2 white
receptacle from the power supply PCB.
7.
Firmly connect the four-pin red and black (DC) cable and connector to its J2 grey
receptacle on the battery interconnect PCB.
8.
Reassemble the monitoring system chassis. Reference
12.7
Front Panel Components Replacement
WARNING:
Before attempting to open or disassemble, disconnect the power cord to
avoid possible injury.
Caution:
Observe ESD (electrostatic discharge) precautions when disassembling and
reassembling the monitoring system and when handling any internal
components.
Caution:
Ensure the work surface is clean and free of debris.
Содержание Nellcor GR101704
Страница 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Страница 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Страница 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Страница 96: ...Operation 6 34 Service Manual a Press an alternate menu option b Press EXIT AND RESTART to exit SERVICE MODE...
Страница 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Страница 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Страница 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Страница 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Страница 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Страница 235: ......