Performance Considerations
8-2
Service Manual
WARNING:
Failure to cover the sensor site with opaque material when operating under
high ambient light conditions may result in inaccurate measurements. Pulse
oximetry readings and pulse signals can be affected by certain environmental
conditions, sensor application errors, and certain patient conditions.
Inaccurate Sensor Measurement Conditions
A variety of conditions can cause inaccurate sensor measurements.
•
Incorrect application of the recommended sensor
•
Placement of the recommended sensor on an extremity with a blood pressure cuff,
arterial catheter, or intravascular line
•
Ambient light
•
Failure to cover the sensor site with opaque material when operating under high
ambient light conditions
•
Excessive patient movement
•
Dark skin pigment
•
Intravascular dyes or externally applied coloring, such as nail polish or pigmented
cream
Signal Loss
Loss-of-pulse signal can occur for several reasons.
•
Recommended sensor applied too tightly
•
Inflation of a blood pressure cuff on the same extremity as the attached
sensor
•
Arterial occlusion proximal to the recommended sensor
•
Poor peripheral perfusion
Recommended Usage
Select an appropriate recommended sensor, apply it as directed, and observe
all warnings and cautions presented in the
Instructions for Use
accompanying
the sensor. Clean and remove any substances such as nail polish from the
application site. Periodically check to ensure the recommended sensor remains
properly positioned on the patient.
Содержание Nellcor GR101704
Страница 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Страница 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Страница 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Страница 96: ...Operation 6 34 Service Manual a Press an alternate menu option b Press EXIT AND RESTART to exit SERVICE MODE...
Страница 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Страница 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Страница 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Страница 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Страница 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Страница 235: ......