![background image](http://html1.mh-extra.com/html/covidien/nellcor-gr101704/nellcor-gr101704_service-manual_2673898089.webp)
Service Mode
Service Manual
6-27
Sound Settings
, p. 10-17, to configure alarm, pulse beep, and
button click volume.
Connectivity Settings
WARNING:
Only use Covidien-approved hardware or remote monitoring software for
data port connectivity.
•
Remote Settings —
Use to set default connectivity for remote monitoring appli-
cations.
•
Nurse Call Polarity —
Use to set default connectivity for nurse call polarity.
•
Serial Connection —
Use to set default connectivity for the serial port.
•
WLAN Settings —
Use to set default connectivity for WLAN networks.
•
LAN Settings —
Use to set default connectivity for LAN networks.
Use the appropriate configuration information to ensure proper connectivity. Any
equipment connected to the data port must be certified according to the latest IEC/
EN 60950-1 standards. All combinations of equipment must be in compliance with
Requirements for Medical Electrical Systems IEC Standard 60601-1: 2007. Anyone
who connects equipment to the data output port is configuring a medical system
and, therefore, is responsible for ensuring the system complies with the Require-
ments for Medical Electrical Systems IEC Standard 60601-1: 2007 and the electro-
magnetic compatibility IEC Standard 60601-1-2: 2007. Accuracy may degrade if it
is connected to secondary I/O devices when equipment is not connected to earth
reference.
•
Connection to a network or data coupling that includes other equipment could
result in previously unidentified risks to patients, operators or third parties, so it is
the responsibility of the person configuring to identify, analyze, evaluate and
control these risks.
•
Any subsequent changes to the network or data coupling, such as network or
data coupling configuration, connection or disconnection of additional or existing
items or equipment, updates or upgrades to items or equipment, might introduce
new risks, so requires re-evaluation and analysis.
When connecting the monitoring system to any other equipment, ensure that
equipment is virus-free. When connecting the monitoring system to equip-
ment to obtain specific patient trend data, verify proper operation of the mon-
itoring system prior to use with a patient. The monitoring system and any
Содержание Nellcor GR101704
Страница 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Страница 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Страница 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Страница 96: ...Operation 6 34 Service Manual a Press an alternate menu option b Press EXIT AND RESTART to exit SERVICE MODE...
Страница 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Страница 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Страница 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Страница 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Страница 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Страница 235: ......