Trend Data Access
7-6
Service Manual
To save specific patient data, firmly connect and properly secure the appropri-
ate cable from the data port to a host system or serial printer. Connect the data
port to a serial printer or host system by using a cable terminated with the fol-
lowing devices.
•
An AMP connector (AMP part number 747538-1)
•
A ferrule (AMP part number 1-747579-2)
•
Compatible pins (AMP part number 66570-2)
The cable should not exceed 25 feet (7.6 meters) in length using RS-232 proto-
col or 4,000 feet (1219.2 meters) in length using RS-422 protocol. The external
ITE (Information Technology Equipment) device must be certified to the latest
IEC/EN Standard 60950-1 standards. The cable used must have a braided shield
that provides 100% coverage. The shield must have a 360-degree connection to
the metal shell on the DB-15 connector and to the connector on the equipment.
No hardware flow control is used. However, support exists for XON/XOFF flow
control in ASCII mode.
100base-T RJ-45 Requirements
Caution:
Do not create sharp bends in the cable. Bend radius cannot exceed one inch.
To save specific patient data, firmly connect and properly secure the appropri-
ate cable from the data receptacle to a receiving device capable of 100base-T
communication. Connect to the data receptacle via an RJ-45 jack properly
attached to a CAT-5 or better cable. The cable should not exceed 100 meters
in length. The external ITE (Information Technology Equipment) device must be
certified to the latest IEC/EN Standard 60950-1 and IEE 802.3 standards.
Note:
The monitoring system also supports 10base-T with a CAT-3 or CAT-4 connection.
Note:
The pin layout illustrates the pins viewed from left to right on the RJ-45 jack,
beginning with pin 8. The tab snap-fits to the receptacle and requires pressure to
release it. The data lines are shielded twisted pair (STP) with an additional drain wire
to reduce crosstalk or noise.
Содержание Nellcor GR101704
Страница 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Страница 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Страница 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Страница 96: ...Operation 6 34 Service Manual a Press an alternate menu option b Press EXIT AND RESTART to exit SERVICE MODE...
Страница 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Страница 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Страница 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Страница 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Страница 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Страница 235: ......