Setting Institutional Defaults
Service Manual
10-7
To import master institutional defaults
Note:
Settings may only be imported from exports created with the same software version.
Do not attempt to import settings from a monitoring system with a different software
version.
Note:
Network settings are not affected by the import.
1.
Access the service function menu. Reference
To access service function menus
2.
Select SETTINGS.
3.
Select IMPORT AND EXPORT SETTINGS.
4.
Select IMPORT SETTINGS.
5.
Insert a USB flash drive containing the master institutional default settings.
6.
Press NEXT.
7.
Press NEXT after ensuring the
ApplicationSettingsFile.xml
option is selected.
8.
Remove the USB flash drive at the prompt.
9.
Press FINISH.
10.
Press EXIT MENU.
11.
Press EXIT AND RESTART.
12.
Press YES.
To restore factory defaults
1.
Access the service function menu. Reference
To access service function menus
2.
Select SETTINGS.
3.
Select RESTORE FACTORY DEFAULTS.
4.
Press YES.
5.
Press FINISH.
Содержание Nellcor GR101704
Страница 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Страница 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Страница 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Страница 96: ...Operation 6 34 Service Manual a Press an alternate menu option b Press EXIT AND RESTART to exit SERVICE MODE...
Страница 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Страница 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Страница 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Страница 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Страница 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Страница 235: ......