Repair
12-16
Service Manual
11.
Remove and set aside the four T-10 torx parameter (interface) board PCB screws.
12.
Gently rotate the parameter board assembly, laying the assembly flat on a static-
free work surface with the oximetry module PCB facing up.
13.
Remove and set aside the two oximetry module PCB T-10 torx screws.
Figure 12-10.
Parameter Board PCB and Oximetry Module Removal
14.
Gently lift upward, removing the oximetry module PCB off the parameter board
assembly standoffs, ensuring all pins fully disconnect from both connectors.
15.
Remove and set aside the failing PCB in an alternate location.
To replace the parameter board assembly
1.
Obtain the replacement parameter board assembly component: the oximetry
module PCB or the parameter board PCB.
2.
Set with the known good component on a static-free work surface.
Содержание Nellcor GR101704
Страница 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Страница 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Страница 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Страница 96: ...Operation 6 34 Service Manual a Press an alternate menu option b Press EXIT AND RESTART to exit SERVICE MODE...
Страница 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Страница 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Страница 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Страница 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Страница 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Страница 235: ......