8-1
8 Performance Considerations
8.1
Overview
This chapter contains information for assisting users to optimize the perfor-
mance of the Nellcor™ Bedside Respiratory Patient Monitoring System.
Prior to initial installation in a clinical setting, have a qualified service technician
verify the performance of the monitoring system per the
Service Manual
.
8.2
Oximetry Considerations
8.2.1
Monitoring System Constraints
•
Pulse Rate —
The monitoring system only displays pulse rates between 20 and
250 bpm. Detected pulse rates above 250 bpm appear as 250. Detected pulse
rates below 20 appear as a zero (0).
•
Saturation —
The monitoring system displays saturation levels between 1% and 100%.
8.2.2
Nellcor™ Sensor Performance Considerations
WARNING:
Incorrect application or inappropriate duration of use of a sensor can cause
tissue damage. Inspect the sensor site as directed in the
Instructions for Use
.
WARNING:
Use only Covidien-approved sensors and interface cables when connecting to
the sensor port. Connecting any other cable or sensor influences the accuracy
of sensor data, which may lead to adverse results.
WARNING:
Pulse oximetry readings and pulse signal can be affected by certain ambient
conditions, sensor application errors, and certain patient conditions.
Содержание Nellcor GR101704
Страница 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Страница 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Страница 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Страница 96: ...Operation 6 34 Service Manual a Press an alternate menu option b Press EXIT AND RESTART to exit SERVICE MODE...
Страница 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Страница 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Страница 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Страница 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Страница 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Страница 235: ......