Repair
12-12
Service Manual
Figure 12-7.
Parameter Module Tab Release
4.
Gently press inward against the parameter module assembly through the two
openings, releasing it from the backplane. Expect minor resistance until it releases.
Figure 12-8.
Parameter Module Assembly Removal
5.
Remove and set aside the parameter module assembly.
6.
Gently rotate the monitoring systemto the upright position.