Modification and Testing
10-18
Service Manual
4.
Press MENU.
5.
Select ALARM LIMITS.
6.
Select SpO
2
.
7.
Slide the SpO
2
Lower Alarm Limit up until it indicates 99. This automatically also
sets the SpO
2
Upper Alarm Limit to 100.
8.
Press SAVE CHANGES.
9.
Press MENU.
10.
Select ALARM LIMITS.
11.
Select PULSE RATE.
12.
Slide the Pulse Rate Lower Alarm Limit up until it indicates 160.
13.
Confirm the following results:
a.
The waveform tracks the pulse rate.
b.
The Pulse Tone is audible.
c.
The SpO
2
and pulse alarms are flashing yellow.
d.
Both audible alarms sound and visual alarms appear, indicating both parame-
ters have violated alarm limits.
Figure 10-2.
Sample: Configuring Alarms
Содержание Nellcor GR101704
Страница 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Страница 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Страница 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Страница 96: ...Operation 6 34 Service Manual a Press an alternate menu option b Press EXIT AND RESTART to exit SERVICE MODE...
Страница 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Страница 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Страница 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Страница 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Страница 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Страница 235: ......