Manufacturer’s Declaration and Guidance
Service Manual
2-13
Enclosure Leakage Current
Condition
AC Line Polarity
Neutral
Line Cord
Power Line
Ground
IEC 60601-1
ANSI/AAMI 60601-1
Normal
Normal
Closed
Closed
100 μA
Single Fault
Open
Closed
500 μA
Closed
Open
Normal
Reversed
Closed
Closed
100 μA
Single Fault
Open
Closed
500 μA
Closed
Open
Table 2-10.
Earth and Enclosure Leakage Current Specifications (Continued)
Table 2-11.
Patient Applied and Patient Isolation Risk Current
Patient Applied Risk Current
Condition
AC Line Polarity
Neutral Line
Power Line
Ground Cable
IEC 60601-1
ANSI/AAMI 60601-1
Normal
Normal
Closed
Closed
100 μA
Single Fault
Open
Closed
500 μA
Closed
Open
Normal
Reversed
Closed
Closed
100 μA
Single Fault
Open
Closed
500 μA
Closed
Open
Patient Isolation Risk Current
Condition
AC Line Polarity
Neutral Line
Power Line
Ground Cable
IEC 60601-1
UL 60601-1
Single Fault
Normal
Closed
Closed
5000 μA
Reversed
Closed
Closed
Содержание Nellcor GR101704
Страница 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Страница 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Страница 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Страница 96: ...Operation 6 34 Service Manual a Press an alternate menu option b Press EXIT AND RESTART to exit SERVICE MODE...
Страница 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Страница 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Страница 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Страница 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Страница 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Страница 235: ......