Repair
12-40
Service Manual
Figure 12-22.
LCD Assembly Removal
8.
Lift out the LCD assembly.
9.
Remove the display flex cable from the LCD assembly.
To replace the LCD Assembly
1.
Replace the new LCD assembly, orienting with the two smaller flex cables to the top.
2.
Thread the large, right-hand flex cable through the plastic LCD clamp plate, ensur-
ing the smaller top flex cable is free above the plastic LCD clamp plate and the
dog-leg flex cable threads through the plastic LCD clamp plate cut-away below
the battery.
3.
Tighten the three T-15 torx screws along the sides of the plate to 10 inch-pounds.
4.
Replace antennae PCBs, ensuring both cables remain unpinched and properly
routed. Reference
Содержание Nellcor GR101704
Страница 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Страница 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Страница 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Страница 96: ...Operation 6 34 Service Manual a Press an alternate menu option b Press EXIT AND RESTART to exit SERVICE MODE...
Страница 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Страница 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Страница 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Страница 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Страница 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Страница 235: ......