Operation
6-28
Service Manual
appropriate equipment must be connected to a grounded AC power source.
Configuring any TCP/IP addresses or entering any network keys requires a USB
keyboard.
To set default remote or serial connectivity
1.
Access the service function menu. Reference
To access service function menus
2.
Select SETTINGS.
3.
Select CONNECTIVITY SETTINGS.
4.
Select the desired option.
a.
Select REMOTE SETTINGS to set the remote connectivity default.
b.
Select SERIAL CONNECTION to set the serial connectivity default.
5.
Select the desired default. Reference
Input and Output Configuration Options
7-2, for a complete listing of setting options.
6.
Select the desired option.
a.
Press SAVE CHANGES to retain the change.
b.
Press CANCEL to leave the default as it was.
To set default WLAN or LAN network connectivity
1.
Access the service function menu. Reference
To access service function menus
2.
Select SETTINGS.
3.
Select CONNECTIVITY SETTINGS.
4.
Select the desired option.
a.
Select WLAN SETTINGS to set the WLAN connectivity default.
b.
Select LAN SETTINGS to set the LAN connectivity default.
5.
Select the desired default. Reference
Input and Output Configuration Options
7-2, for a complete listing of setting options. Reference
p. 7-9, particularly in reference to the icon table for the configuration interface.
a.
Select WI-FI to set a wireless connectivity default, using a USB keyboard to
input any TCP/IP address or network key.
b.
Select NETWORK CONNECTION to set the LAN connectivity default, using a
USB keyboard to input any TCP/IP address or network key.
6.
Press FINISH to retain the change.
Содержание Nellcor GR101704
Страница 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Страница 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Страница 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Страница 96: ...Operation 6 34 Service Manual a Press an alternate menu option b Press EXIT AND RESTART to exit SERVICE MODE...
Страница 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Страница 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Страница 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Страница 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Страница 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Страница 235: ......