Power Components Replacement
Service Manual
12-31
12.6.4
Left Power Cable Assembly Replacement
To remove the left power cable assembly
1.
Disassemble the monitoring system chassis. Reference
2.
Move the front chassis off to the side.
3.
Identify the proper cables under the battery cradle. They are the cables to the
opposite side of the enclosure from the power entry module.
4.
Remove the five-pin Mini-mate black cable and connector from its J2 receptacle
from the power supply PCB.
5.
Remove the four-pin red and black (DC) cable and connector with ferrite bead
from its J2 receptacle on the battery interconnect PCB.
Figure 12-18.
Left Power Cable Assembly Replacement
Содержание Nellcor GR101704
Страница 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Страница 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Страница 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Страница 96: ...Operation 6 34 Service Manual a Press an alternate menu option b Press EXIT AND RESTART to exit SERVICE MODE...
Страница 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Страница 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Страница 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Страница 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Страница 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Страница 235: ......