Service Mode
Service Manual
6-23
4.
Select SENSOR ALARM PRIORITIES.
5.
Select the desired priority (LOW, MEDIUM, or HIGH) for each of the sensor alarms.
6.
Press SAVE CHANGES.
7.
Press EXIT MENU to return to the Service Main Menu.
Display Settings
Provides means of creating and transferring institutional display settings.
•
Monitoring Layout —
Use to set the default monitoring screen layout. The
factory default is pleth only.
•
Main Screen Trend Parameters —
Use to set the default trends for the main
monitoring screen trend information. Default is SpO
2
and pulse.
•
Main Screen Trend Scale —
Use to set the default trend scale when viewing
trend information. Default is one (1) hour.
•
Monitoring History Scale —
Use to set the default trend scale when viewing
the monitoring history. Default is one (1) hour.
•
Backlight Brightness —
Use to set the default backlight brightness. Default is
75% bright.
•
Allow Backlight OFF —
Use to set permission for turning backlight completely
off. Default is NO.
•
Wake Display on Alarm —
Use to set the default for waking the display during
an alarm condition. Default is YES.
•
Language —
Use to set the default interface language. Default is English.
To set the default monitoring screen layout
1.
Access the service function menu. Reference
To access service function menus
2.
Select SETTINGS.
3.
Select DISPLAY SETTINGS.
4.
Select MONITORING LAYOUT.
5.
Select the desired default.
Содержание Nellcor GR101704
Страница 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Страница 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Страница 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Страница 96: ...Operation 6 34 Service Manual a Press an alternate menu option b Press EXIT AND RESTART to exit SERVICE MODE...
Страница 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Страница 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Страница 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Страница 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Страница 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Страница 235: ......