Operation
6-22
Service Manual
To set the alarm limit threshold defaults
1.
Access the service function menu. Reference
To access service function menus
2.
Select SETTINGS.
3.
Select ALARM LIMITS.
4.
Select the desired mode: ADULT or NEONATE.
Note:
Pulse rate and SpO
2
thresholds are different for adults and neonates. In addition,
the SPD alarm management parameter is not available for neonates.
5.
Select the desired type of limit.
a.
Select SpO
2
and slide bars up or down to set upper and lower alarm limits.
b.
Select PULSE and slide bars up or down to set upper and lower alarm limits.
c.
Select SATSECONDS and set to 10, 25, 50, or 100 SatSeconds, or OFF.
d.
Select SPD and set the sensitivity to 1, 2, or 3, or OFF.
e.
Select PULSE RATE DELAY to 5 or 10 seconds, or OFF.
6.
Press SAVE CHANGES.
7.
Press EXIT MENU to return to the Service Main Menu.
Sensor Alarm Priorities
Institutional default alarm priorities can be set for the following sensor alarms:
•
Sensor Disconnected
•
Sensor Off
•
Sensor Failure
By default, these sensor alarms are low priority. They can be configured inde-
pendently as high, medium, or low.
To set sensor alarm priorities
1.
Access the service function menu. Reference
To access service function menus
2.
Select SETTINGS.
3.
Select ALARM SYSTEM.
Содержание Nellcor GR101704
Страница 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Страница 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Страница 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Страница 96: ...Operation 6 34 Service Manual a Press an alternate menu option b Press EXIT AND RESTART to exit SERVICE MODE...
Страница 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Страница 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Страница 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Страница 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Страница 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Страница 235: ......