Repair Prerequisites and Required Equipment
Service Manual
12-5
Note:
Some spare parts come with an enclosed business reply card. After receiving the spare
parts, please complete and return the business reply card.
12.3
Repair Prerequisites and Required Equipment
WARNING:
Only qualified service personnel should open the chassis, remove and replace
components, or make adjustments. If your medical facility does not have a
qualified service technician, please contact Covidien Technical Services or
your local Covidien representative.
WARNING:
Use the test data sheet to ensure the monitoring system passes all safety,
performance, and functional tests prior to use in a clinical setting.
WARNING:
Before attempting to open or disassemble, disconnect the power cord to
avoid possible injury.
Caution:
Observe electrostatic discharge (ESD) precautions prior to opening the chassis
or handling any internal components.
Caution:
Observe the required torque for tightening screws. Over-tightening can strip
out screw holes, rendering them useless.
Note:
Do not dispose of parts by placing in the regular trash. Dispose of parts in accordance
with local guidelines and regulations or contact Covidien to arrange for disposal.
Only a qualified service technician may disassemble the monitoring system to
its major component parts. The supported replacement level is to the printed
circuit board (PCB) and major subassembly level. After isolating the problem to
a suspected PCB, follow the procedures for disassembly, then replace the orig-
inal suspect PCB with a known good PCB. Verify the symptom disappears and
Содержание Nellcor GR101704
Страница 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Страница 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Страница 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Страница 96: ...Operation 6 34 Service Manual a Press an alternate menu option b Press EXIT AND RESTART to exit SERVICE MODE...
Страница 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Страница 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Страница 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Страница 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Страница 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Страница 235: ......