![background image](http://html1.mh-extra.com/html/covidien/nellcor-gr101704/nellcor-gr101704_service-manual_2673898198.webp)
Repair
12-10
Service Manual
12.5
Chassis Disassembly and Reassembly
WARNING:
Before attempting to open or disassemble, disconnect the power cord to
avoid possible injury.
Caution:
Observe ESD (electrostatic discharge) precautions when disassembling and
reassembling the monitoring system and when handling any internal
components.
Caution:
Ensure the work surface is clean and free of debris.
Note:
The battery charge procedure should be performed before repairs when possible.
Disassembly and reassembly of the chassis requires removal of the parameter
module. Prepare a reasonable, static-free work surface for opening and
working on both halves. The following section describes three primary tasks.
1.
, p. 12-10, to remove and then replace
the parameter module or its subassemblies.
2.
Monitoring System Chassis Disassembly
, p. 12-17, for chassis disassembly.
3.
Monitoring System Chassis Reassembly
, p. 12-20, for chassis reassembly.
12.5.1
Parameter Module Replacement
Remove the parameter module prior to performing any internal repair to the
monitoring system. The parameter module consists of the parameter module
case and the parameter board assembly. The following subsection describes
three primary tasks.
1.
, p. 12-10, to remove and then replace
the parameter module assembly.
2.
Parameter Module Case Replacements
, p. 12-13, to remove and then
replace components of the parameter module case.
3.
Parameter Board PCBA Replacements
, p. 12-15, to remove and then replace
either the parameter board PCB or the OEM PCB.
Содержание Nellcor GR101704
Страница 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Страница 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Страница 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Страница 96: ...Operation 6 34 Service Manual a Press an alternate menu option b Press EXIT AND RESTART to exit SERVICE MODE...
Страница 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Страница 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Страница 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Страница 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Страница 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Страница 235: ......